Table 3-6. Manometer Troubleshooting
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
1. NO PRESSURE READING.
Pressure is not being applied to either, or both sidesof
Step 1.
the instrument.
Check for plugged or leaking pressure lines. Make
applicable corrections.
Step 2. Atmospheric pressure connection not vented to
atmosphere.
Check atmosphere pressure connection and ensure proper
venting.
Foreign particles in internal passages of the meter are
Step 3.
blocking the flow.
Remove mercury from meter and check internal passages
for foreign particles. Clean instrument as required.
2. HIGH OR LOW PRESSURE READINGS.
Step 1. Check for leaks or obstructions.
Check lines for leaks or obstructions. Make applicable
corrections,
Step 2. Check if tubing and/or mercury may contain foreign
particle matter.
Disassemble and clean instrument. Replace mercury.
Step 3. Check for improper zero setting.
Check zero setting.
Make proper adjustment.