TM 5-6640-213-14
(3) Remove and replace the defective motor coupling; make
sure that the ejector blades are to the left and are one-quarter of
an inch above the edge of the mold.
Adjust the bin-arm switch as
Adjusting the Bin-Arm Switch.
follows ..
(1) Remove the front cover.
(2) Slide the bin-arm switch to the right or left, as
needed, until the switch clicks off and back on when the arm is
within 1 inch of the freezing mechanism.
A troubleshooting chart for the ice maker
h.  Troubleshooting.
is listed in table 4-2.
CLASS S WEIGHTS.
4-6.
The class S weights must be certified at 720-day intervals by
the A-level calibration facility.  The weight must be certified
within the following tolerances.
Maximum
Minimum
Gram Weights
99.99975 grams
100.00025 grams
100 grams
49.99988 grams
50.00012 grams
50 grams
19.999926 grams
20.000074 grams
20 grams
10.000074 grams
9.999926 grams
10 grams
4.999926 grams
5.000054 grams
5 grams
2.000054 grams
2 grams
1.999946 grams
1.000054 grams
0.999946 grams
1 gram
REID VAPOR PRESSURE APPARTUS.
4-7.
a.  Before Operation Service.  Insure that the overflow pipe is
tight.  Fill bath with water up to level of overflow pipe.
A solution of copper sulfate in water is poisonous.
NOTE
The position of the overflow pipe is set to main-
tain the level of the water at least l-inch above
the immersed bomb, as specified by ASTM.  Add 1/2
gram of copper sulphate to the bath to create a
solution of 10 ppm to minimize algae growth.