TM 5-6640-213-14
DIRECT SUPPORT AND GENERAL SUPPORT MAINTENANCE INSTRUCTIONS
Section I.
REPAIR PARTS, SPECIAL TOOLS AND EQUIPMENT
There are no special tools or equipment required for maintenance
of the laboratory.
4-2.
MAINTENANCE REPAIR PARTS.
Repair parts and equipment covering direct and general support
maintenance of the laboratory are listed in Appendix C.
TROUBLESHOOTING
Section II.
a. This section contains repair, adjusting and troubleshooting
information and tests for locating and correcting most troubles which
may develop for items which require direct support and general
support maintenance.
b. This manual cannot list all possible malfunctions that may
occur, or all tests or inspections and corrective actions. If a
malfunction is not listed, or is not corrected by listed corrective
actions, you should notify higher lever maintenance.
4-4.
OVEN.
a. Compound Adjustment, Repair, or Replacement. Before making
adjustment or repairs, or before replacing components, disconnect the
unit from it electrical source and remove the shelves and thermometer.
(1) Oven door. If heat loss occurs, inspect the door
closure. The plate on the door that engages the magnetic catch on
the cabinet is adjustable. Loosen or tighten it, as applicable, to
assure proper door closure. Also inspoect the door gasket to make
certain it fits firmly against the cabinet at all points. Replace
the gasket, if damaged.
(2) Pilot light. Access to the back side of the control
panel is gained by removing the two top and two bottom screws that
hold. it in place. When replacement of the pilot lights is necessary
tag the lead wires with respect to their relative positions, detach
them from the light, and push it outward from the back of the panel.
Press the new light into this opening and attach the proper wires to
it. Be sure the red portion is on the right side.
|