TM 5-6640-213-14
the macro weight counter and set the zero as described in paragraph
(4) above. Then, recheck the sensitivity,If required, readjust the
sensitivity, set the zero, and recheck. Continue the sequence until
the balance is correctly adjusted.
NOTE
Whenever the tare boat is to be used, the tare ring
is to be removed. The boat and ring have identical
weight.
If the beam is not arrested when the transportation
arresting device is removed, serious damage to the
balance may result.
Uncrating and Preparing the Balance for Use.
b.
(1) Uncrating. The analytical balance is packed in the
original carton and strapped to the multilane shock-absorber support.
The multilane support is the normal work surface on which the balance
rests.
(a) Cut away the straps and carefully open the carton
at the top.
(b) Lift the balance out of the carton and place it
on the support. Remove the plastic shroud and set it aside, but do
not destroy it.
(c) Remove the accessory carton from the weighing
compartment after removing the strips of tape on the sliding doors.
The accessory carton contains the weighing pan (1, fig. 3-7), stirrup
bearing plate (2), chamois leather and camel's hair brush. The for-
mer items must be installed in the balance; the latter should be
stored for use in cleaning.
3-27
|