|
|
TM 5-6640-213-14
(e) and (f) for the alarm set
(g)
Repeat
steps
(d),
point.
The air-purging system is located
(5) Air-Purging System.
It consists of a grill and exhaust
in a cabinet opposite the sink.
through wall mounted louver assembly and out of the laboratory at the
rate of 600 cubic feet per minute (cfm).
NOTE
The
air-purging system is also activated
when
the main power switch is turned on
and
it can be operated manually by the
fan
switch beneath the light switch.
Test
Equipment.
f.
(1)
Double
Beam
Balance.
The double beam balance (NSN 6670-00-
(a) Description.
436-9857) is supplied for weighing chemicals that are to be used in
various testing procedures.
The upper beam is graduated from O to 10
The lower beam is graduated from O to
grams in O.1-gram divisions.
Two removable, 6-inch (15.2 cm) dia-
200 grams in 10-gram divisions.
The capa-
meter, stainless steel pans are supplied with the balance.
city of the balance without the use of loose weights is 210 grams.
Since the capacity of the balance exceeds the beam capacity, the use
of additional loose weights is necessary to take advantage of the
A set of 12 brass, tall-form, loose weights (NSN 6670-
full capacity.
00-803-9680) is included with the laboratory equipment for use with
The weights are positioned in a vinyl plastic block and
the balance.
they range in size from 1 to 1,000 grams.
(b)
Operating
Procedures.
1.
Set each poise at the zero mark.
Always check balance zero before each use. If
2.
the instrument does not balance at exactly zero, make the necessary
corrections with the two knurled nuts on the front of the beam.
When the balance" has been restored, lock the nuts against each other.
Place the specimen to be weighed on the left
3.
platform
of
the
balance.
Move the lower poise to the right until the
4.
Move
first notch is reached; this causes the right platform to drop.
the poise back one notch; this should cause the right platform to rise
again.
2-17
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |